浩瀚色谱建立聚乙二醇1中环氧乙烷和二氧六环的测定方法.方法:采用HH-1毛细管柱(30 m×0.53 mm×3.00μm),程序升温,氢火焰离子化检测器.结果:环氧乙烷、二氧六环分别在0.61~ 30.32μg·mL-1、5.85~292.6 μg·mL-1浓度范围内与峰面积的线性关系良好.环氧乙烷高、中、低三浓度水平平均回收率为106.9%,其RSD为3.5%(n=9);二氧六环高、中、低三浓度水平平均回收率为107.2%,其RSD为2.2%(n=9).结论:建立的测定方法灵敏、准确,可用于聚乙二醇中环氧乙烷与二氧六环的测定.
GB 1886.302-2021 食品安全国家标准 食品添加剂 聚乙二醇
本标准适用于环氧乙烷和水缩聚而制得的食品添加剂聚乙二醇。
环氧乙烷和二氧六环的测定
名称:毛细管柱
型号:HH-1
规格:30m*0.32mm*0.5um
*高温度:300°C
编号:20240322002
应用:GB 1886.302-2021 食品安全国家标准 食品添加剂 聚乙二醇
Establishment of a method for the determination of ethylene oxide and dioxane in polyethylene glycol 1 using vast chromatography. Method: HH-1 capillary column (30 m x 0.53 mm x 3.00) was used μ m) Program heating, hydrogen flame ionization detector. Results: Ethylene oxide and dioxane were found to be between 0.61 and 30.32, respectively μ G · mL-1, 5.85~292.6 μ The linear relationship between the peak area and the concentration range of g · mL-1 is good. The average recovery rate of ethylene oxide at high, medium, and low concentration levels is 106.9%, with an RSD of 3.5% (n=9); The average recovery rate of high, medium, and low concentrations of dioxane is 107.2%, with an RSD of 2.2% (n=9). Conclusion: The established determination method is sensitive and accurate, and can be used for the determination of ethylene oxide and dioxane in polyethylene glycol
GB 1886.302-2021 National Food Safety Standard Food Additive Polyethylene Glycol
This standard applies to the food additive polyethylene glycol produced by the condensation of ethylene oxide and water.
Determination of ethylene oxide and dioxane
Name: Capillary Column
Model: HH-1
Specification: 30m * 0.32mm * 0.5um
*High temperature: 300 ° C
Number: 20240322002
Application: GB 1886.302-2021 National Food Safety Standard Food Additive Polyethylene Glycol